Monitor AMI-II CACE Degasser AC

Specifications

Process measuring device for conductivity before / after cation exchange and after degassing, including continuous resin regeneration.

Complete system with:

  • Transmitter AMI-II CACE in a rugged aluminum enclosure (IP 66).
  • Three Swansensors UP-CON1000 SL conductivity sensors with slot-lock design and integrated Pt1000 temperature sensor, k = 0.04 cm-1.
  • Flow cells Catcon-Plus SL CACE with digital sample flow meter and B-Flow UP-CON SL made of stainless steel 316L. Quick sensor release with patented slot-lock design.
  • EDI-Module with exchangeable sample chamber module and automatic deaeration.
  • Thermal degasser with minimal power consumption and sample flow.


Conductivity measurement  
Measuring range 0.055 to 1000 µS/cm
Resolution 0.001 µS/cm
Accuracy (at 25 °C) ±1 % of measured value or ±1 digit (whichever is greater).
Response time (t90, specific conductivity) < 5 sec.
pH and alkalizing reagent calculation  
Range pH (25 °C) 7.5 to 11.5
Range concentration (e.g. ammonia) 0.01 to 10 ppm
Sample conditions  
Flow rate 5 to 6 L/h
Temperature up to 50 °C
Inlet pressure (25 °C) 0.5 bar
Outlet pressure pressure free
EDI capacity max. SC = 40 µS/cm as NH4OH
max. SC = 350 µS/cm as NaOH
Power supply  
Voltage 100 - 240 VAC (± 10 %)
50/60 Hz (± 5 %)
Power consumption max. 180 VA


For more information, please refer to the technical data sheet.

IXA23582100 AMI-II CACE DG_Tech.png