Monitor AMI Hydrogen QED AC
Specifications
Dissolved Hydrogen measurement
Measuring range
0,1 to 800 ppb or 0 to 50% saturation
Resolution
0.01 ppb (up to 9.99 ppb)
0.1 ppb (from 10.0 to 99.9 ppb)
1 ppb (from 100 to 800 ppb)
Accuracy
+- 5% of reading or +- 0,5 ppb
(whichever is greater)
Repeatability
+- 1% of reading or +- 0,5 ppb
(whichever is greater)
Temperature measurement
Measuring range
-30 to 130°C
Resolution
0.1° C
Power Supply
Voltage
100-240 VAC (+- 10%), 50/60 Hz (+- 5%) or
10-36 VDC (+- 10%)
Sample conditions
Flow rate
6 to 20 l/h
Temperature
up to 45° C
Inlet pressure
up to 1 bar
Outlet pressure
pressure free
Suspended solids
less than 10 ppm
Dimensions
Panel
280 x 850 x 150 mm
For further information see datasheet.