Monitor AMI Deltacon Degasser AC

Description

Triple channel conductivity monitor for specific (total), acid (cation) and degassed conductivity.

  • Calculation of the pH value and alkalizing reagent concentration in power plants by differential conductivity measurement (VGB-S-010-T-00).
  • Complete system with dedicated transmitter, three high precision conductivity sensors and two flow cells; one flow cell with integrated cation column, needle valve and digital sample flow meter.
  • Sample reboiler unit with heating and cooling system made of stainless steel.
  • DG electronic controller for sample reboiler unit with vapor pressure control (IP66)
  • Simple exchange of resin catridge; recommended for use with Swan rinsed nuclear grade resin for optional performance.
  • Factory tested on wet bench. Ready for installation and start-up.
IXA23481100_AMI_DELTACON_DG_oG