Monitor AMI LineTOC AC

Specifications

Process instrument for the reagent-free measurement of the TOC of ultrapure water by means of UV oxidation. Complete system with transmitter, UV reactor, two high accuracy conductivity sensors, sample flow detection and device for calibration and verification on 400 x 850 mm stainless steel mounting plate.

TOC Measurement


Measuring range:

0.00 to 1000 ppb TOC

Resolution:

0.01 ppb

Reproducibility:
0,1 to 50 ppb
50 to 1'000 ppb


± 1 ppb
± 2%

Accuracy conductivity:
0.055 to 2 µS/cm (20 °C)


±1 %

Response time

<2 min

 

Sample conditions


Flow rate:

1 to 5 L/h

Temperature:

10 to 40 °C
with sample cooler: up to 90°C

Inlet pressureAbs. (25 °C):

up to 1.5 bar
with pressure regulator: up to 5 bar

Outlet pressure:

pressure free

Conductivity:

0.055 to 2 µS/cm

Particle size:

<100 µm
No sand, no oil

 

Power supply


Voltage:

100 – 240 VAC (±10 %),

Power consumption:

50/60 Hz (±5 %)
max. 55 W


For more information, please refer to the technical data sheet.


IXA23612100_AMI_LineTOC.jpg