Monitor AMI-II LineTOC compact version AC

Specifications

Specifications

Process instrument for the reagent-free measurement of the TOC of ultrapure water by means of UV oxidation.
Complete system with transmitter, UV reactor, two high accuracy conductivity sensors, sample flow meter and device for calibration and verification on 700 x 450 mm stainless steel mounting plate.



TOC Measurement

Measuring Range:0.00 to 1000 ppb TOC
Resolution: 0.01 ppb
Reproducibibility: 0.1 to 50 ppb ±1 ppb
                             50 to 1000 ppb ±2 %Accuracy Conductivity 0.055 to
                             2 µS/cm (25 °C) ±1 ppb



Sample Conditions

Flow rate 3 to 6 L/h

Temperature 10 to 40 °C / with sample cooler: up to 90 °C

Inlet Pressure (Abs.) up to 1.5 bar / with pressure regulator: up to 5 bar, 80 °C

Outlet Pressure pressure free

Conductivity 0.055 to 2 µS/cm

Particle Size<100 µm / no sand no oil



Power Supply

AC Version: 100 - 240 VAC (± 10 %)
                    50/60 Hz (± 5 %)
Power Consumption: max. 55 VA



Dimensions

180 x 142 x 94mm



For more information, please refer to the technical data sheet.

 

IXA23513100 AMI-II LineTOC_compact