Specifications
Process measuring device for conductivity before
and after cation exchange with continuous resin regeneration.
Complete system with:
- Transmitter AMI-II CACE in
a rugged aluminum enclosure (IP 66).
- Two Swansensors UP-CON1000 SL conductivity sensors with slot-lock design and integrated
Pt1000 temperature sensor, k = 0.04 cm-1.
- Flow cell Catcon-Plus SL CACE made of stainless steel 316L with digital sample flow meter.
Quick sensor release with patented slot-lock design.
- EDI-Module with
exchangeable sample chamber module and automatic deaeration.
| Conductivity measurement |
|
| Measuring
range |
0.055 to 1000 µS/cm |
| Resolution |
0.001 µS/cm |
| Accuracy
(at 25 °C) |
±1 % of measured value or ±1
digit (whichever is greater). |
| Response
time (t90, specific conductivity) |
< 5 sec. |
| pH and
alkalizing reagent calculation |
|
| Range pH
(25 °C) |
7.5 to 11.5 |
| Range
concentration (e.g. ammonia) |
0.01 to 10 ppm |
| Sample
conditions |
|
| Flow
rate |
3 to 4 L/h |
| Temperature |
up to 50 °C |
| Inlet
pressure (25 °C) |
0.5 bar |
| Outlet
pressure |
pressure free |
| EDI
capacity |
max. SC = 40 µS/cm as
NH4OH
max. SC = 350 µS/cm as NaOH |
| Power
supply |
|
| Voltage |
100 - 240 VAC (± 10 %)
50/60 Hz (± 5 %) |
| Power
consumption |
max. 35 VA |
For
more information, please refer to the technical data sheet.